ATLAS PATIENT MONITOR - POWER SUPPLY BOARD - FUSE LOCATION'S AND MORE [MEDICAL EQUIPMENT]



POWER SUPPLY REMOVAL
  • Slide the power supply out from the E-Pac Foam.
  • Power Supply Fuse Location
NOTE: Notice power supply and power supply fuse location.
  • Fan Removal
Slide fan out from E-Pac foam.
  • E-Pac Removal (Lower)
Remove lower E-Pac foam from unit.
  • Deflection PCB Removal
Carefully remove the small deflection PCB from the CRT.
  • CRT Removal
Remove the two machine screws holding CRT to the front panel.
NOTE: Notice how the ground wire is attached and is located along the side of CRT.
NOTE: The CRT and deflection PCB are factory tuned as a unit and are sold only together as a unit.
  • Display PCB Removal
Remove all 9 self-tapping screws that hold the Display PCB to the Bezel.
NOTE: Notice the correct orientation of CRT screw and washer.
  • Printer Door Button Removal
Use a small screwdriver to push the door button tab up and away from the pin that locks it onto the printer frame.
NOTE: Notice how the printer button is being removed.
  • Printer Assembly Locking Tab Release.
Pull the locking tab down and towards the front of the rear housing assembly.
  • Printer Positioning Slot Removal
Make sure printer door is open. Begin to pull the positioning slots towards the front of the rear housing and then out from the rear housing.
  • Printer Assembly Removal.
After printer positioning slots have cleared the front rear housing, pull the printer down and out of the rear housing.
  • Printer Lip Assembly.
Observe how the printer assembly fits into the lip of the rear housing.
  • Printer Cable Routing.
Observe how the printer cable lies over the printer PCB.