Part-3
END.
Principles of operation to Terumo Infusion Pump Model-508 follows on next blog
27 |
|
|
|
28 |
|
|
|
31 | 1. maximum discharge pressure adjusting mechanism defective. | 1. If delivery limit on test is less than 98ml or more than 103ml, adjustment is improper. | 1. Re-adjust pressure. Compensate delivery rate. (replace seal packing, if degenerate) |
33 | 1 connector (CN-8),external connection connector defective. 2 Cable defective. 3 External alarm relay output circuit defective. 4 External alarm relay defective. 5 Main CPU defective. |
|
|
34 |
|
|
|
35 |
|
| 1.Check connector function.
|
37 |
| 1. If occlusion pressure is less than 0.5kg/cm2 or more than 1.0kg/cm2 | 1. Adjust occlusion detection pressure. |
38 |
|
| 1 Replace main PCB. Replace main PCB. |
39 | 1. Main Cpu defective. | 1 Source power shutdown does not take place. Confirm voltage A/D value in the sef diagnosis mode. If source power shutdown does not take place even A/D value is les than 101, main CPU is defective. | 1. Replace main PCB. For the method of changing the mode into self diagnosis and confirming voltage A/D value, refer paragraph 4.2 |
40 |
|
|
|
41 |
|
|
|
42 |
|
|
|
END.
Principles of operation to Terumo Infusion Pump Model-508 follows on next blog