TROUBLESHOOTING
Static Manometer Accuracy Test
BP Leak Test
Static Manometer Accuracy Test
Static Manometer Accuracy Test
Required materials.
- Serial interface cable
- PC with HyperTerminal (Windows 95 or above)
- 500cc vessel with inflation system.
- Calibrated manometer
- Turn computer on and activate Hyper terminal.
- 9600 Baud, 8 bit word, 1 stop bit no parity, no flow control ANSI character set Start àPrograms àAccessories àCommunication àHyperTerminal
- Turn Atlas on.
- Enter the following command sat the PANGEA> prompt.
PANGEA> bp valve close<ENTER>
Pressurize the Atlas monitor to the applied
pressures in Table.
- To compare the applied pressure to the transducer readings use the following command.
- PANGEA> bp press<ENTER>
- Verify that the primary transducer and secondary transducer are within the specifications listed in Table.
NOTE: The Atlas monitor will return a response to the
computer screen as follows.
If you tested at the 50mmHg pressure you would receive a
message similar to: px = 5279,sx = 5305 x 0.01. Where px is the primary
transducer and 5279 = 52.79 mmHg and sx is the secondary transducer and 5305 =
53.05 mmHg.
NOTE: Unit must not leak more than 5 mmHg in a 15 second
interval while attached to a 100cc test cavity pressurized at50 mmHg, 150 mm Hg
and 250 mmHg.
- Required Materials: 100cc test vessel (+10cc/-0cc), stopwatch, and squeeze bulb, calibrated manometer.
- Connect Atlas to 100cc test vessel and calibrated manometer.
- Turn Atlas on.
- Press DATE/TIME button then press LEAD SELECT button to access the Service Mode.
- Press SELECT button and choose Verify manometer.
- Pressurize the Atlas with squeeze bulb to 50mmHg and allow reading to stabilize.
- Observe pressure for 15 seconds.
- Unit should not leak more than 5mmHg during this time.
- Repeat this process at 150mmHg and 200mmHg.
BP Dump Verification Test
- Connect the Atlas to a 500 cc vessel.
- Turn Atlas on.
- Press DATE/TIME button then press LEAD SELECT button to access the Service Mode.
- Select Verify manometer.
- Pressurize the Atlas to just over 260 mmHg.
- Allow reading to stabilize for 15 seconds.
- Ensure the reading is still over 260 mmHg.
- Start the stop watch when you press the SELECT button. When the SELECT button is pressed the Atlas will start a pressure dump.
- Verify that the pressure drops to below 15 mmHg within 10 seconds.
Hardware Fail Safe Tests
- NOTE: Atlas hardware must detect over pressure on unit pneumatic system between 296.0 mmHg and 330.0 mmHg.
- Connect the Atlas for testing.
- Turn Atlas on.
- Press CLOCK/DATE button then press LEAD SELECT button to access Service Mode.
- Press SET button to highlight the Manometer.
- Pump system up and record at what pressure the Atlas dumps.
ECG/Respiration Test
NOTE: Simulator should support ECG impedance respiration.
- Cycle power on Atlas, connect a 5 lead ECG cable set to the simulator. Configure the simulator for NSR, 100 bpm and set the impedance respiration rate to 20.
- Press DATE/TIME button then press TREND button to access Advanced Configuration menu.
- Press SELECT button to highlight ECG lead set and press SET button to select 5 wire.
- Set ECG gain to Automatic, ECG speed to 25mm/s, ECG bandwidth to Monitor, Second trace selection to ECG.
- Press TREND button to return to idle screen.
- Verify you see:
2. Scale bar on top left of CRT
3. Heart rate displayed as set on simulator
4.Lead Selected = II
5. Pulse tone high pitched- Press LEAD Select button and step through each of the lead settings. (I,II,III,aVR,aVL,aVF,V)
- Verify that you see a different ECG waveform for each lead selected. Heart Rate will go to dashes and alarms on some leads.
- Set lead selected to II.
- Press DATE/Time button then press Trend button to access Advanced Configuration menu.
- Change Second trace selection to Respiration. Press Trend button to return to idle screen.
- Verify that you see:
2. Scale bar on left of top line
3. Heart rate displayed as set on simulator
4. Lead Selected = II 5. Pulse tone high pitched
6. Respiration trace on second line and rate as set on
simulator
7. Respiration rate displayed on CRT as set on simulator.
- Disconnect ECG simulator.
SpO2 Test
- Turn SpO2 alarms on.
- Connect SpO2 cable to simulator.
- Set simulator pulse rate to 60 beats/minute and verify that pulse rate is 60 ±3 beats/minute or3%, whichever is greater.
- Set SpO2 between 70% and 100% and verify that SpO2 is accurate ±2 digits.