Sunday, March 24, 2013

PRECOR TREADMILL - FAULT DISPLAY CODES - REMEDIES - 1 of 2



FAULT DISPLAYS                   [1 of 2]


Error 00, 01, 02, 03, 04 - Memory, Ram & EEPROM Test Errors

Description: Errors 00 through Error 04 check upper PCA memory locations, RAM memory locations and EEPROM checksum during the power up test sequence. If a fault is found during the power up test sequence, the appropriate error code will be displayed.
Possible Causes
·        These errors messages almost always indicate an upper PCA problem when they are consistently displayed. Failures causing these error messages to be displayed are rare.
Possible Remedies
·        If the error message is consistently displayed when the unit is powered up, the upper PCA should be replaced. It is possible to see one of these messages displayed intermittently due to external causes. If the units A.C. input significantly dips during the power up test sequence the test could fail resulting in one of these messages being displayed. Treadmills operating  on non-dedicated A.C. circuits may see these messages displayed on an intermittent basis

Error 05 - Key Depressed at Power Up
Description: The power up test sequence has detected a key in the operated condition. The power up test sequence is performed every time the unit is turned on (powered up). This test is performed because a permanently stuck key will inhibit the correct operation of the unit.
Possible Causes
·        A key on the display housing is permanently stuck in the operated condition.
·        One of the upper PCA’s keypad interface chips has failed.
·        Liquid (perhaps perspiration) is present on the surface of a touch sensitive display.
Possible Remedies
·        Remove the keypad connector cable from the upper PCA. Restart the power up sequence, if the ERROR 5 is no longer present, the keypad is the cause. The keypad is typically part of the display housing. Replace the display housing to correct the problem.
·        Remove the keypad connector cable from the upper PCA. Restart the power up sequence, if the ERROR 5 is still present, the upper PCA is the cause. Replace the upper PCA to correct this problem. Most upper PCA’s are provided without software. Therefore, the software (chip) must be removed from the upper PCA being replaced and re-installed on the replacement upper PCA.
·        Wipe the surface of the touch sensitive display clean of all liquid. Restart the power up sequence, if the ERROR 5 is no longer present, the liquid was the cause.

Error 09 - Lower PCA Memory Test Errors
Description: Error 09 checks lower PCA memory locations during the power up test sequence. If a fault is found during the power up test sequence, error 09 will be displayed. This error code is only used on products utilizing the dual microprocessor system.
Possible Causes
·        This error message almost always indicates a lower PCA problem when it is consistently displayed. Failures causing this error message to be displayed are rare.
Possible Remedies
·        If the error message is consistently displayed when the unit is powered up, the lower PCA should be replaced. It is possible to see this error message displayed intermittently due to external causes. If the units A.C. input significantly dips during the power up test sequence the test could fail resulting in this message being displayed. Treadmills operating on non-dedicated A.C. circuits may see this message displayed on an intermittent basis.

Error 10 - Line Frequency out of Acceptable Range
Description: This error is present on treadmills using an SCR motor controller and the C544 EFX. Because the treadmill running belt speed (and C544 resistance) is derived from the A.C. line frequency, it is critical that the unit’s operating system can correctly identify the A.C. line as either 50 Hz. or 60 Hz. and ensure that the line frequency is within acceptable tolerances.
Possible Causes
·        The treadmill is operating on a non-dedicated A.C. circuit.
·        On 120 V.A.C. circuits the A.C. hot and A.C. neutral wires are reversed.
·        Incorrect jumper setting on the upper PCA
·        The fuse feeding the treadmills stepdown transformer is blown.
·        The A.C. line frequency is incorrect.
·        The A.C. line may be extremely noisy.
Possible Remedies
·        Treadmills must be installed on a completely dedicated 20 amp A.C. circuit. Both the hot and neutral leads must be dedicated to the treadmill. If another piece of equipment is sharing a circuit with the treadmill it can create enough electrical noise to make A.C. line frequency identification impossible. The treadmill’s A.C. circuit must be reconfigured as a completely dedicated 20 amp circuit.
·        Most line frequency detection systems monitor the A.C. line frequency on the hot A.C. line. In the case of a reversed 120 V.A.C circuit, the monitoring is taking place on the neutral (ground) side of the A.C. line. Many of the monitoring systems are incapable of detecting line frequency on the neutral wire. It must first be determined if the reversal is in the A.C. wiring feeding the treadmill or in the treadmill internal wiring. The hot and neutral wires can be verified by measuring each in reference to A.C. (green wire) ground. The hot lead will read full line voltage (typically 105 to 120 V.A.C.) and the neutral line should read approx. 0 V.A.C.
·        Some upper PCA’s had jumper setting for use with “SCR” or “PWM” drive circuits. If the upper PCA jumper was set for “PWM” operation on an SCR unit, error 10 would be displayed.  Change the upper PCA jumper setting to “PWM” to correct the problem.
·        Some treadmills (C944 PWM version 2 and C96X PWM version 2) utilized a pair of fuses feeding the stepdown transformer for the low voltage power supply on the lower PCA. The A.C. line frequency was monitored on the output of one of the fuses. If the fuse blew, A.C. line frequency could no longer be detected even though the treadmill appeared normal otherwise.
·        Even though unlikely, the A.C. line frequency could actually be out of acceptable limits. This is more likely to occur in countries where A.C. power systems may not be well developed and controlled. There may be little that can be done about this condition. If the source frequency can not be corrected the only possibility may be replacing the treadmill with a treadmill utilizing a PWM motor controller. PWM motor controllers do not depend on line frequency for speed control.
·        Sometimes A.C. being fed into the distribution system is too electrically noisy to allow correct line frequency identification. This can be caused by other equipment within the A.C. distribution system creating the electrical noise. Identifying the source of the electrical noise can be difficult. There may be little that can be done about this condition. If the source can not be located or the source can not be removed or made electrically quiet the only possibility may be replacing the treadmill with a treadmill utilizing a PWM motor controller. PWM motor controllers do not depend on line frequency for speed control.

Error 11 - Watchdog (Upper PCA)
Description:  The error 11 watchdog monitors the upper PCA low voltage power. If the low voltage power drops below a preset level, the error 11 will be displayed.
Possible Causes
·        A failure in the upper PCA overloads the low voltage power supply and causes it to drop below the preset limit.
·        A fault in the interconnect cable (lower PCA to upper PCA) causing the input voltage to the upper PCA to be too low.
Possible Remedies
·        If the error message is consistently displayed when the unit is powered up, the upper PCA should be replaced. It is possible to see this message displayed intermittently due to external causes. If the units A.C. input significantly dips during the power up test sequence the test could fail resulting in this message being displayed. Treadmills operating on non-dedicated A.C. circuits may see this message displayed on an intermittent basis.
·        Substitute a known good interconnect cable in place of the existing cable to determine if the interconnect cable is the cause. For the purpose of the test, the substitute cable should be connected directly between the upper and lower PCA’s.

Error 12 - Watchdog (Lower PCA)
Description:  The error 12 watchdog monitors the lower PCA low voltage power. If the low voltage power drops below a preset level, the error 12 will be displayed.
Possible Causes
·        A failure in the lower PCA overloads the low voltage power supply and causes it to drop below the preset limit.
·        A fault in the interconnect cable (lower PCA to upper PCA) causing the lower PCA low voltage power supply be too low.
·        A failure in the upper PCA that overloads the lower PCA low voltage power supply causing it to be too low.
Possible Remedies
·        If the error message is consistently displayed when the unit is powered up, the lower PCA should be replaced. It is possible to see this message displayed intermittently due to external causes. If the units A.C. input significantly dips during the power up test sequence the test could fail resulting in this message being displayed. Treadmills operating on non-dedicated A.C. circuits may see this message displayed on an intermittent basis.
·        Substitute a known good interconnect cable in place of the existing cable to determine if the interconnect cable is the cause. For the purpose of the test, the substitute cable should be connected directly between the upper and lower PCA’s.
·        Substitute a known good upper PCA for the existing PCA to determine if the upper PCA is the cause

Error 13 - Fan at Incorrect Speed
Description:  This error monitored the rotating speed of an external cooling fan attached to the drive motor on 9.55, C954 version 1 and C956 version 1 treadmills. The system used an optical speed sensor that read an infrared signal reflected back to the speed sensor from a target attached to the hub of the cooling fan.
Possible Causes
·        It was determined that the fan speed sensing system was unreliable and false error 13 indications were being displayed. The lower software was changed to remove all error 13 indications.
·        If the installed software includes an error 13, the fan may be clogged with dust and/or dirt. 
Possible Remedies
·        Replace the lower PCA software with the most current lower PCA software available.
·        Clean the fan. If the error 13 still occurs, replace the lower PCA software with the most current lower PCA software available.

Error 20 - Too many Maximum Power Requests in 1 Second Description This error monitors the treadmills power bit reading when the treadmill was operating at speeds of 1 mile per hour or less. Power bits indicate the relative time that the motor controller drive circuit is turned on. Therefore power bits are an indication of the amount of power the load is demanding from the motor controller. Error 20 indicates that at a speed of 1 mile per hour or less an excessive amount of power is being demanded from the motor controller for an excessive period of time.
Possible Causes
·        The treadmill user (especially a heavier user) is standing on the drive belt as they start the running belt into motion. Starting the running belt into a very heavy load can cause the error 20 condition.
·        The running bed and/or running belt is badly worn and the error 20 occurs when the user steps onto the running belt.
·        The drive motor brushes are badly worn and not making sufficient contact with the drive motor commutator. The error 20 occurs when the user steps onto the running belt. The treadmill may also bog down when the user steps onto the running belt.
·        On treadmills utilizing an SCR motor controller, a leaky or shorted drive motor filter capacitor could raise the current demand sufficiently to cause the error 20.
·        On treadmills utilizing an SCR motor controller, a leaky or shorted drive motor inductor could raise the current demand sufficiently to cause the error 20.
A shorted or de-magnetized drive motor is causing the high power demand.
Possible Remedies
·        The user should be instructed to straddle the running belt and start the treadmill at a low speed (1 mile per hour or less) and then step onto the running belt.
·        A clamp-on A.C. ammeter must be used to determine the amount of A.C input current being drawn by the treadmill under no load and loaded conditions. The current readings can be used to determine the condition of the running belt and/or running bed. These same readings should be taken on a similar unit with a known good running belt and running bed. Those readings can then be used as a benchmark for future use. Replace the running belt and/or running bed as required. Remember, many running beds are double sided and the bed can be flipped over rather than being replaced.
·        Replace both drive motor brushes.
·        A clamp-on ammeter should be used to measure the A.C. input current at no load. The filter capacitor should then be electrically isolated from the drive motor circuit. The drive motor leads must be kept intact so that the drive motor will operate with the filter capacitor removed from the circuit. The no load A.C. input current should be measured again, if the A.C. input current is significantly lower, replace the filter capacitor.
·        Use the same procedure as in step 4 above, except remove and isolate the inductor from the drive motor circuit instead of the filter capacitor.
·        A known good drive motor must be substituted for the existing drive motor.

Error 21 - Too many Maximum Consecutive Power Requests
Description:  This error monitors the treadmills power bit reading when the treadmill was operating at speeds above 1 mile per hour. Power bits indicate the relative time that the motor controller drive circuit is turned on. Therefore power bits are an indication of the amount of power the load is demanding from the motor controller. Error 21 indicates that at a speed of greater than 1 mile per hour an excessive amount of power is being demanded from the motor controller.
Possible Causes
·        The running bed and/or running belt is badly worn and the error 21 occurs while the treadmill is being used.
·        The drive motor brushes are badly worn and not making sufficient contact with the drive motor commutator. The error 21 occurs while the treadmill is being used. The treadmill may also bog down when the load or speed is increased.
·        Some lower PCAs require a jumper when installed in 120 V.A.C. treadmills.
·        On treadmills utilizing an SCR motor controller, a leaky or shorted drive motor filter capacitor could raise the current demand sufficiently to cause the error 21.
·        On treadmills utilizing an SCR motor controller,a leaky or shorted drive motor inductor could raise the current demand sufficiently to cause the error 21.
·        A shorted or de-magnetized drive motor is causing the high power demand.
Possible Remedies
·        A clamp-on A.C. ammeter must be used to determine the amount of A.C input current being drawn by the treadmill under no load and loaded conditions. The current readings can be used to determine the condition of the running belt and/or running bed. These same readings should be taken on a similar unit with a known good running belt and running bed. Those readings can then be used as a benchmark for future use. Replace the running belt and/or running bed as required. Remember, many running beds are double sided and the bed can be flipped over rather than being replaced.
·        Replace both drive motor brushes.
·        If the lower PCA requires the 120 VAC jumper and the jumper is not installed, the treadmill will typically display an error 21 between 7 and 8 M.P.H. Install the lower PCA jumper to correct this problem. WARNING:Be sure the treadmill is a 120 V.A.C. unit before installing the jumper. Installing the jumper on a 240 V.A.C. will cause the lower PCA to fail.
·        A clamp-on ammeter should be used to measure the A.C. input current at no load. The filter capacitor should then be electrically isolated from the drive motor circuit. The drive motor leads must be kept intact so that the drive motor will operate with the filter capacitor removed from the circuit. The no load A.C. input current should be measured again, if the A.C. input current is significantly lower,replace the filter capacitor.
·        Use the same procedure as in step 4 above, except remove and isolate the inductor from the drive motor circuit instead of the filter capacitor.
·        A known good drive motor must be substituted for the existing drive motor

Error 22 - No Motor Pulses at Start Up
Description:  Error 22 indicates that the drive motor (running belt) has been instructed to start, by either manual or program control, and the monitoring system has not received any response from the speed sensing system indicating that the drive motor has started.
Possible Causes
·        The treadmill utilizes an optical (infrared) speed sensor and the line of sight between the speed sensor and the speed sensor target is obstructed.
·        Drive motor brushes are bad.
·        The speed sensor position relative to the speed sensor target is incorrect.
·        The speed sensor potentiometer adjustment is incorrect (when equipped).
·        The speed sensor is inoperative.
·        The lower PCA is bad.
·        The lower PCA to upper PCA interconnect cable is bad.
·        The upper PCA is bad
Possible Remedies
·        Carefully, vacuum out the drive motor compartment with a static safe vacuum cleaner. Pay special attention to the area between the speed sensor and speed sensor target.
·        If the motor brushes do not make sufficient contact with the motor commutator, the drive motor will not start when drive voltage is applied. Replace both drive motor brushes.
·        Adjust the speed sensor position so that the space between the speed sensor and the speed sensor target is 3/16 inch and so that the speed sensor photo eye is aimed at the larger portion of the targets wedge shaped section. On hall effect speed sensors (magnetic sensor) the space between the end of the speed sensor and the flywheel tooth should be 3/16 inch.
·        Locate the speed sensor potentiometer on the lower PCA. Set the speed sensor potentiometer at mid-range. Start the treadmill and increase the speed to 6 mile per hour. Carefully adjust the speed sensor potentiometer to the point where the treadmill is operating the smoothest. Increase the treadmills speed to full speed at repeat the previous adjustment. These two adjustments are typically small and are usually only a matter of fine tuning.
·        If the speed sensor in inoperative the drive motor will start and run briefly before the error 21 occurs. If it is possible to access the speed sensor wiring, measure the D.C. voltage between the speed sensors leads Two of the leads will constantly read approximately 5 Vdc. Two of the leads will toggle between 0 Vdc. and 5 Vdc. as you slowly rotate the speed sensor target. If you do not see both measurements above, the speed sensor is inoperative and must be replaced. In some cases the speed sensor is mounted on the lower PCA and the the lower PCA must be replaced.
·        Substitute a known good lower PCA to determine if the lower PCA is bad.
·        Substitute a known good interconnect cable for the existing cable.
·        Substitute a known good upper PCA to determine if the upper PCA is bad.

Error 23 - Motor Pulses Missing after Start Up
Description:  Error 23 indicates that the speed sensor signal was lost while the treadmill was in operation.
Possible Causes
·        The treadmill utilizes an optical (infrared) speed sensor and the line of sight between the speed sensor and the speed sensor target is obstructed.
·        The speed sensor position relative to the speed sensor target is incorrect.
·        The speed sensor potentiometer adjustment is incorrect.
·        The speed sensor or drive motor wiring has an intermittent connection.
·        The drive motor brushes are badly worn and not making sufficient contact with the drive motor commutator
·        The lower PCA is bad.
·        The lower PCA to upper PCA interconnect cable is intermittent.
·        The upper PCA is bad
Possible Remedies
·        Carefully, vacuum out the drive motor compartment with a static safe vacuum cleaner. Pay special attention to the area between the speed sensor and speed sensor target.
·        Adjust the speed sensor position so that the space between the speed sensor and the sped sensor target is 3/16 inch and so that the speed sensor photo eye is aimed at the larger portion of the targets wedge shaped section
·        Locate the speed sensor potentiometer on the lower PCA. Set the speed sensor potentiometer at mid-range. Start the treadmill and increase the speed to 6 mile per hour.  Carefully adjust the speed sensor potentiometer to the point where the treadmill is operating the smoothest. Increase the treadmills speed to full speed at repeat the previous adjustment. These two adjustments are typically small and are usually only a matter of fine-tuning.
·        Physically test all wiring and wire connections associated with the speed sensor and drive motor. Replace or repair wiring and wire connections as appropriate.
·        Replace both drive motor brushes.
·        Substitute a known good lower PCA to determine if the lower PCA is bad.
·        Substitute a known good interconnect cable for the existing cable.
·        Substitute a known good upper PCA to determine if the upper PCA is bad.

Error 24 - Reduce Speed Requested, Speed is not Reducing
Description: Error 24 indicates that the the speed control system has instructed the treadmill to slow down and the response from the speed sensing systemindicates that the treadmill has not slowed. This error will typically occur when the actual operating speed is greater than the speed being requested under either manual or program control.
Possible Causes
·        The treadmill user (especially a heavier user) is operating the treadmill at a relatively a low speed and a high incline. Some treadmills, under the above conditions, will operate at a speed slightly higher than the selected speed because of the combination of very low running bed and running belt friction and the downward force of the users weight.
·        The lower PCA is bad.
Possible Remedies
·        The user must be informed that this error can occur under low speed, high incline conditions. They must also be aware that the treadmill speed control system must test for an over speed condition. The test and resulting error condition is for their protection.
·        The typical lower PCA failure mode producing an error 24, causes the drive motor to start immediately upon power up (without a manual or program drive motor start request). Replace the lower PCA.

Error 25 - Motor Pulse Count over Acceptable Limit
Description:  This error monitors the treadmills power bit reading. Power bits indicate the relative time that the motor controller drive circuit is turned on. Therefore power bits is an indication of the amount of power the load is demanding from the motor controller. The treadmill software sets a maximum allowable power bit limit. Error 25 indicates that the maximum power bit limit is being exceeded.
Possible Causes
·        The running belt and/or running bed is badly worn.
·        On treadmills utilizing an SCR motor controller, a leaky or shorted drive motor filter capacitor could raise the current demand sufficiently to cause the error 25.
·        On treadmills utilizing an SCR motor controller,a leaky or shorted drive motor inductor could raise the current demand sufficiently to cause the error 25.
·        A shorted or de-magnetized drive motor is causing the high power demand.
Possible Remedies
·        A clamp-on A.C. ammeter must be used to determine the amount of A.C input current being drawn by the treadmill under no load and loaded conditions. The current readings can be used to determine the condition of the running belt and/or running bed. These same readings should be taken on a similar unit with a known good running belt and running bed. Those readings can then be used as a benchmark for future use. Replace the running belt and/or running bed as required. Remember, many running beds are double sided and the bed can be flipped over rather than being replaced.
·        A clamp-on ammeter should be used to measure the A.C. input current at no load. The filter capacitor should then be electrically isolated from the drive motor circuit. The drive motor leads must be kept intact so that the drive motor will operate with the filter capacitor removed from the circuit. The no load A.C. input current should be measured again, if the A.C. input
          current is significantly lower,replace the filter capacitor.
·        Use the same procedure as in step 4 above, except remove and isolate the inductor from the drive motor circuit instead of the filter capacitor.
·        A known good drive motor must be substituted for the existing drive motor.

Error 26 - Motor Pulse Width Incorrect
Description:  This error monitors the speed sensor signal and verifies the speed sensor signal is appropriate for the requested speed. If the speed sensor signal is incorrect or erratic an Error 26 will be displayed.
Possible Causes
·        The treadmill utilizes an optical (infrared) speed sensor and the line of sight between the speed sensor and the speed sensor target is obstructed.
·        The speed sensor position relative to the speed sensor target is incorrect.
·        The speed sensor potentiometer adjustment is incorrect (when equipped).
·        The speed sensor target is damaged.
Possible Remedies
·        Carefully, vacuum out the drive motor compartment with a static safe vacuum cleaner. Pay special attention to the area between the speed sensor and speed sensor target.
·        Adjust the speed sensor position so that the space between the speed sensor and the speed sensor target is 3/16 inch and so that the speed sensor photo eye is aimed at the larger portion of the targets wedge shaped section. On hall effect speed sensors (magnetic sensor) the space between the end of the speed sensor and the flywheel tooth should be 3/16 inch.
·        If the treadmill utilizes a speed sensor potentiometer, locate the speed sensor potentiometer on the lower PCA. Set the speed sensor potentiometer at mid-range. Start the treadmill and increase the speed to 6 mile per hour. Carefully adjust the speed sensor potentiometer to the point where the treadmill is operating the smoothest. Increase the treadmills speed to full speed at repeat the previous adjustment. These two adjustments are typically small and are usually only a matter of fine-tuning.
·        Most speed sensor targets consist of a circular arrangement of alternate black and silver sections. If the speed sensor target is torn or badly scratched the speed sensor may receive an incorrect reading. Replace the speed sensor target.

Error 27 - Too Much Drive Motor Current
Description:  Some treadmill motor controllers monitor the amount of current being delivered to the drive motor. The software sets a maximum amount of allowable drive motor current. This error indicates that maximum drive motor current has been reached.
Possible Causes
·        The running belt and/or running bed is badly worn.
·        A shorted or de-magnetized drive motor is causing the high power demand.
Possible Remedies
·        A clamp-on A.C. ammeter must be used to determine the amount of A.C input current being drawn by the treadmill under no load and loaded conditions. The current readings can be used to determine the condition of the running belt and/or running bed. These same readings should be taken on a similar unit with a known good running belt and running bed. Those readings can then be used as a benchmark for future use. Replace the running belt and/or running bed as required. Remember, many running beds are double sided and the bed can be flipped over rather than being replaced.
·        A known good drive motor must be substituted for the existing drive motor.

Error 28 - Temperature Too High
Description:  Some treadmill motor controllers monitor the temperature of the motor controller output switching device. Typically, these motor controllers use a fan to force cool the output devices heat sink. This error indicates that the heat sink temperature has exceeded maximum.
Possible Causes
·        The lower PCA cooling fan is clogged.
·        The lower PCA cooling fan is inoperative.
·        The running belt and/or running bed is badly worn.
·        The lower PCA is defective.
·        A shorted or de-magnetized drive motor is causing the high power demand
Possible Remedies
·        Check the cooling fan mounted on the lower PCA to ensure that the fan is not clogged with dust. Thoroughly clean the fan and ensure that it spins freely. Some lower PCA’s do not have a cooling fan and some lower PCA’s have the fan mounted on the underside of the the lower PCA and are not immediately visible.
·        The lower PCA is clean and unobstructed but the fan does not spin. Ensure that the fan wiring is securely and correctly connected to the lower PCA. If the fan is unobstructed and the fan wiring is good, replace the lower PCA.
·        A clamp-on A.C. ammeter must be used to determine the amount of A.C input current being drawn by the treadmill under no load and loaded conditions. The current readings can be used to determine the condition of the running belt and/or running bed. These same readings should be taken on a similar unit with a known good running belt and running bed. Those readings can then be used as a benchmark for future use. Replace the running belt and/or running bed as required. Remember, many running beds are double sided and the bed can be flipped over rather than being replaced.
·        Substitute a known good lower PCA to determine if the lower PCA is defective.
·        Substitute a known good drive motor to determine if the drive motor is defective

Error 30, 31, 32 and 33 - Communications Error
Description:  Errors 30 through 33 all indicate the loss of communications or erratic communications between the microprocessors in the upper PCA and the lower PCA. The trouble shooting procedures for the entire are essentially the same. Error 30 is defined as upper PCA not receiving from lower PCA. Error 31 is defined as faulty data received from lower PCA. Error 32 is defined as lower
PCA not receiving from upper PCA. Error 33 is defined as faulty data received from upper PCA
Possible Causes
·        An upper PCA to lower PCA interconnect cable is bad.
·        The interconnect cable at the upper PCA was mistakenly plugged into the CSAFE connector.
·        A 5.17 EFX equipped with version 1.01 upper and lower software.
·        A 5.17 with mismatched upper and lower software.
·        A defective lower PCA.
·        A defective upper PCA.
Possible Remedies
·        Substitute a know good interconnect cable between the upper and lower PCAs to determine if the interconnect cable is defective.
·        Remove the interconnect cable from the CSAFE connector and insert it in the correct upper PCA connector.
·        Replace the 5.17 upper and lower software with the most current software versions.
·        If only one of the upper or lower software is version 1.01, replace the version 1.01 software with the most current version of software.
·        Substitute a known good lower PCA to determine if the lower PCA is defective.
·        Substitute a known good upper PCA determine if the upper PCAs defective.