In some cases a panel may have defective cells that cannot be controlled. These defective cells can be categorized into three types:
- Non - Lighting Cell Defect: Defect in which the cell is always Off.
- Non-Extinguishing Cell Defect: Defect in which the Cells are always On.
- Flickering Cell Defect: Defect in which the cell is flickering.
Measuring area
Cell defect specification
ITEM
|
|
SPECIFICATION
|
|
Number of cell defects
|
Distance between cell defects
|
Non-lighting cell defect
|
Zone-A: 2 and less
Zone – B: 8 or less
|
Regardless of A and B zone
|
Non-Exitithgushing Cell defect
|
Zone – A: Zero
Zone-B: 2 and less
|
-Distance between NLs is 15mm
-Distance between NL and NEF is 15mm
-Distance between NEF is 30mm
|
Flickering cell defect
|
Zona-A: Zero
Zone-B: 2 and less
|
|
Total defect
|
Total number of cell defects in Zone A and Zone B is 8 and
less
|
Note:
NL = Non Lighting Cell defect.
NEF = Non Extinguishing effect.
And Flickering
cell defect.
|